For a printable version of this instruction sheet, click
here.
*Print Version Requires Adobe Acrobat Reader.
Click the button above for a FREE download of Acrobat Reader.
|
|
|
|
SABRATEK PUMP INSTRUCTIONS
I. PUMP SETUP
| A. |
Pump setup with primary IV administrator set. |
| |
- Switch the pump on by pressing the ON/OFF KEY. The words
"SELF-TEST
" will appear briefly, followed
by the name of the current IV administration setting ("ABBOT
SET!", "BAXTER SET!", "McGAW SET!",
"SABRASET!", etc.). This is done in order to verify
that the correct set is going to be used with the infusion.
WARNING! USE ONLY THE IV ADMINISTRATION
SET INDICATED ON THE DISPLAY. CHANGE SETTING TO SABRASET.
- Prepare the IV solution container and prime the IV administration
set according to the instructions provided on the set package.
Make sure there is no air in the set. Close the administration
set clamp.
- Hang the IV solution container so the container's bottom
is approximately 14" to 20" above the pump.
- Open the OUTER DOOR of the pump by pushing on the red
label marked PUSH TO OPEN located on the right-hand side
of the OUTER DOOR LATCH. Swing the OUTER DOOR LATCH down.
- On the lower, right-hand corner of the inner door, press
the blue label marked PUSH TO LOAD to open the FREE FLOW
CLAMP. Slip the IV administration set tubing UNDER the FREE
FLOW CLAMP, in the directions of the flow direction arrows,
and position the tubing in the TUBING GUIDES. Release the
pressure on the INNER DOOR. The tubing will be firmly clamped
in place without permitting any free flow in the IV set.
- Swing the OUTER DOOR up and push firmly on the OUTER DOOR
until the DOOR LATCH clicks into place.
- Open the IV administration set's clamp and attach the
end of the IV set to the patient's access device.
|
II. PROGRAMMING THE PUMP
| A. |
Continuous |
| |
- After selecting the correct pump set, press YES.
- The word "PROGRAMMABLE" will be displayed. Press
YES.
- The words "PRESSURE:HI" or "PRESSURE:LO"
(showing the current pressure alarm setting) will be displayed
briefly followed by the word "CONTINUOUS?".
- The words "DELAY DLVRY?" will appear. Press
YES.
- You can now choose whether or not to delay the start of
the infusion:
If you want to delay delivery:
Press YES. The word "Delay: 00:00" will appear.
Enter the delay time (in hours:minutes) and press YES when
the desired
delay time is displayed.
If you don't want to delay delivery:
Press NO. The word "RATE?" will be displayed.
- You can now choose between entering a delivery rate or
a delivery time (the delivery time is the actual time of
the infusion, excluding any delay time):
If you want to enter an infusion rate:
Press YES. "Rate: XXX" will be displayed. If the
displayed rate is correct,
press YES to accept it.
To change the displayed rate, enter the desired infusion
rate and press YES when the correct rate is displayed.
If you want to enter an infusion time:
Press NO. The word "TIME?" will appear.
Press YES and "TIME: XX:XX" will be displayed.
If the displayed time is correct, press YES to accept it.
To change the displayed time, enter the desired infusion
time (in hours:minutes) and press YES when the correct time
is displayed.
The pump will display "VOLUME: XXXX".
- If the displayed volume is correct, press YES to accept
it. To change the displayed volume, enter the desired volume
and press YES when the correct volume is displayed.
- The words "Bag Vol:" will appear alternating
with either the word "Total:" or the word "Rate:".
(If you previously entered a delivery rate, then the corresponding
time will be shown as "Total:"; if you previously
entered a delivery time, then the corresponding "Rate:"
will be shown.) This gives you an opportunity to confirm
all infusion parameters. Press NO to check or change the
parameters.
- Press YES. The words "READY CONT" will be displayed.
All parameters have been accepted by the pump and will remain
in the pump's memory until they are changed. The delay delivery
value will be available only for the first time that this
program is run. On all subsequent continuous deliveries,
the program will immediately begin, unless a delay delivery
is reprogrammed into the profile.
- If the programmed infusion is for immediate delivery,
press YES or RUN/HOLD when the words "READY COUNT"
are displayed. The infusion rate and the volume-to-be-infused
will be displayed, alternating with the word "CONTINUOUS".
This provides you with an opportunity to reconfirm the infusion
parameters.
- Press RUN/HOLD and the infusion will start. While infusing,
the pump will display the infusion rate and the volume to
be infused, alternating every 15 seconds with the word "CONTINUOUS".
- If the programmed infusion is intended for a later delivery,
switch the pump off by pressing ON-OFF. The programmed parameters
will remain in the pump's memory until they are changed.
|
| B. |
Intermittent
This profile permits the delivery of a number of doses at set
intervals (intermittently) over a period of time up to 48 hours
with a preset KVO rate between each dose. You determine the
infusion's total delivery time (up to 48 hours), number of doses,
dose volume, dose time or dose rate, and KVO rate. Programming
this profile is
possible only if the pump is in the Programmable mode. |
| |
- Switch the pump on by pressing ON/OFF. The words "SELF-TEST
"
followed by the name of the IV administration set that must
be used with the pump will be displayed ("ABBOTT SET!",
"BAXTER SET!", "McGAW SET!", "SABRASET!",
etc.).
- Press YES. The word "PROGRAMMABLE" will be displayed.
- Press YES. the words "PRESSURE:HI" or "PRESSURE:LO"
(showing the current pressure alarm setting) will be displayed
briefly followed by the word "CONTINUOUS?".
- Press NO. The words "AUTO-RAMP?" will be shown.
- Press NO. The word "INTERMITTENT?" will be displayed.
- Press YES. The words "DELAY DLVRY?" will display.
- You can now choose whether or not you wish to delay the
start of the delivery:
If you want to delay delivery:
Press YES. The word "Delay: 00:00" will appear.
Enter the delay time (in hours:minutes) and press YES when
the desired
delay time is displayed.
If you don't want to delay delivery:
Press NO.
- The word "Total:24hrs?" will be displayed.
you can now choose the total time period that you want the
doses to be delivered over.
If you want a 24 hour delivery period:
Press the YES key.
If you want a specific delivery period:
Press the No key. "OTHER TOTAL?" will display.
Press YES. "Total: XX:XX will display. If the displayed
time is correct, press YES to accept it.
To change the displayed time, enter the total delivery period
that you want
(in hours:minutes) and press YES when the correct time is
displayed.
- "# Doses: X" will appear. If the displayed number
of doses is correct, press YES to accept it. To change the
displayed doses, enter the total number of doses that you
wish to deliver and press YES when the correct number of
doses are displayed.
- The words "DOSE TIME?" will be displayed.
- You can now choose between entering a delivery time or
a delivery rate for a single dose (all of the doses will
have the same delivery time or the same delivery rate):
If you want to enter a dose time:
Press YES. "DosTime: XX:XX" will be displayed.
If the displayed time is correct, press YES to accept it.
To change the displayed time, enter the dose time you want
(in hours:minutes) and press YES when the correct delivery
time of the dose is displayed.
If you want to enter a dose rate:
Press NO. The words "DOSE RATE?" will be displayed.
Press YES. "DosRate: XXX" will appear. If the
displayed rate is correct, press YES to accept it.
To change the displayed rate, enter the dose rate and press
YES when the correct rate is displayed. "Dos Vol: XXXX"
will appear. This is the volume that will be delivered for
each dose.
- If the displayed dose volume is correct, press YES to
accept it. To change the displayed dose volume, enter the
dose volume you want and press YES when the correct volume
is displayed. "KVO Rate: X" will be displayed.
This is the rate at which the pump will deliver between
the doses and during the delay delivery time (if you chose
to delay the start of the infusion).
- If the displayed KVO rate is correct, press YES to accept
it. To change the displayed KVO rate, enter the desired
KVO rate (1-9 ml/hr) and press YES when the correct KVO
rate is displayed.
- The words "Bag Vol:" alternating with "DosTime:"
or "DosRate:" will be shown (if you had entered
a dose time, the corresponding "DosRate:" will
be shown; if you had entered a dose rate, the corresponding
"DosTime:" will be shown). This gives you an opportunity
to confirm the infusion parameters. Press NO to
check or change the parameters.
NOTE: You must enter a KVO rate that
is less than the dose rate. If the entered KVO rate is higher
than the dose rate, the pump will not accept the entered
parameter.
- Press YES. The words "READY INTERM" will be
displayed. All parameters have been accepted by the pump
and will remain in the pump's memory until they are changed.
The delay delivery value will be available only for the
first time that this program is run. On all subsequent intermittent
deliveries, the program will immediately begin with Dose
1, unless a delay delivery is reprogrammed into the profile.
- If the programmed infusion is for immediate delivery,
press YES or RUN/HOLD. The dose rate and the total volume-to-be-infused
will be displayed, alternating with the work "INTERMITTENT".
This provides you with an opportunity to reconfirm the infusion
parameters.
- Press RUN/HOLD and the infusion will start. While infusing,
the pump will display the current infusion rate, as well
as the cumulative volume infused. This information will
alternate every 15 seconds with the words "DOSE 1",
"KVO DOSE 1", "DOSE 2", "KVO DOSE
2", etc. (depending on which phase of the infusion
regimen the pump is presently delivering).
- If the programmed infusion is intended for a later delivery,
switch the pump off by pressing ON/OFF.
|
For any questions or concerns, call your
home IV nurse according to the instructions your nurse has given
you. Horizon Healthcare Services can also be reached 24 hours a
day at (717) 544-3590 or toll free at (800) 848-3204
|
|